Table 2: Error Codes and Troubleshooting
Problem Cause(s) Solution(s)
Err 0 This typically indicates that the
platform is too light.
Ensure that the stainless steel cover is properly in
place. Turn the scale off and remove the platform.
Replace the platform and reapply power. If error
persists, the weight sensor or circuit board may be
faulty and you should contact your distributor.
Err 1 Overweight error. The sample
is too heavy for the balance.
Remove some of the sample. If sample is within the
balance capacity, balance may need to be calibrated.
Err 2 Underweight error. This error
is similar to Err 0 as it
indicates that the platform is
not properly installed.
Power off the balance and re-seat the platform before
reapplying power to the balance. The balance may
need to be calibrated.
Err 3 Non-zero error. This indicates
that there was weight on the
balance during power up.
Remove any weight(s) from the platform, turn the
balance off and back on.
F1 AZT A key was pressed during
power up.
Turn the balance off and back on again.
F2 Cnt A key was pressed during
power up.
Turn the balance off and back on again.
Incorrect Weight
Reading
Balance was not set to zero or
was not calibrated.
Press Tare to zero the balance. Calibrate the balance
as necessary.
No Display Power adapter not properly
connected or batteries are
dead.
Reconnect power adapter or replace batteries.